Calibration Settings

Before calibration, you need to determine the axis direction of calibration. The set calibration axis determines the upward axis (Z or Y) in the Global Coordinate System after calibration. Set the type of L-bar for calibration according to the length of the L-bar you are using (if you are unsure about the type of L-bar, you can consult our technical engineers).

Click on "Calibration Pane" in the toolbar at the top of the window. In the calibration window, click the gear icon. Under the "Frame" option, confirm the needed coordinate system definition. You can choose whether the coordinate system is Z-axis up or Y-axis up through the "Up Axis" . In the "L-Frame Type" option, select the corresponding L-Type. The L-Type is related to the purchased L-Type rod. The size type of the L-Type rod can be determined by measuring the distance between the two Marker points at the farthest end of the long side of the L-Type rod. For example, if you measure the distance between the two Marker points at the farthest end of the long side of the L-Type rod to be 600mm or nearly, then the type of L-Frame you are using is "L-600". You can also customize the L-bar type. Select "Manual" from the dropdown box. Customization requires you to manually enter the X, Y, and Z coordinates of each marker ball in the "Coordinates" table below.

Last updated