AD Analog Access to AMTI Force Plates

Accessing AMTI force plates via AD requires the use of camera trigger synchronization. Connect one end of the synchronization cable to the AUX port at the back of the camera with IP 201, and the other end to the NI AD acquisition card through an adapter and jumper cable. According to the names like Fx, Fy on the AMTI analog signal cable, correctly connect the analog signal cables to the NI AD acquisition card as per the wiring requirements of the AMTI force plate;

Driver Installation:

  • Ensure the signal amplifier is connected to the force plate and the amplifier is powered on.

  • In the Control Panel, search for "Device Manager" and click it.

  • Find "Other devices," which is marked with a yellow warning (16.3.1).

  • Right-click and select "Update Driver", choose "Browse my computer for driver software". Using AMTI's USB drive, find the "64-bit Drivers Ver. 3.4.7" folder in the following path on the USB. Select this folder and click next . After the progress bar ends, "Windows has successfully updated your drivers" will be displayed, indicating that the driver installation is complete.

  • If you don't have a driver folder, please contact the AMTI manufacturer to get it.

Install "NetForce":

  • For the installation of "NetForce" software, please contact our engineering technical staff to copy or send the "NetForce Ver software installation package". Click "setup" to start the installation and follow the installation prompts to complete the software installation.

Configure the matrix in "NetForce" software:

  • Open the "NetForce" software, click "setup" at the top of the software, and select "Hardware Installation" (16.3.2). In the "Hardware Installation" window, select "Optima \Gen-5" from the dropdown list under "New Installation" (16.3.3), and click "Install".

  • In the pop-up "AMTI System Configuration" window, select "Fully Conditioned" from the dropdown list under "Analog Outputs" (16.3.4); click the Optima button at the bottom of the window, click Continue, and click Browse. In the subsequent window (16.3.5), navigate to the path and select the "10382M.bcf" file (16.3.6), click Open and wait for the download to finish. Click Done to exit once the download is complete.

  • In the Installed Amplifier section of the AMTI System Configuration window, click the Configure button underneath. In the pop-up window, set all Excitation values of Fx/Fy/Fz/Mx/My/Mz to 10 and set all Gain values to 1000. Click the Analog Adjust button at the bottom, and record all six values in the Analog Scale Factor row (16.3.7). Divide 1000 by each value in the Analog Scale Factor to get the matrix value in the XINGYING software. Click Apply all, then OK to close the window.

  • In the AMTI System Configuration window, click Apply then Save to preserve this configuration for future use, then click Done to exit. When exiting the NetForce software, you will be prompted whether to save these settings (16.3.8); click "Yes" to automatically close the software. Once the NetForce software is closed, unplug the USB cable connecting the AMTI force plate amplifier to your computer.

Open XINGYINM -- click on the dynamometer menu bar -- select Force Platform Configuration, then click the "+" to add a force plate. On the right, set the parameters for the AMTI force plate. Starting from the top left to the bottom right of the calibration matrix, enter the matrix values derived from the "AMTI NetForce" software, then click apply.

Click on the dynamometer menu bar -- select Analog Configuration. In the Analog Configuration window, configure the analog channels for AMTI. Right-click in the list and select "AMTI--FP1". Configure the range for each channel according to the output range of the test platform. Each analog channel can select ranges like +/-10mV, +/-10V, etc., from the "Range" drop-down menu. Click the "Select" button at the top left of the window to select all analog channels.

In the Force Platform Configuration window, check the box for Analog Access (16.3.9), connect the camera, play, and you can use the AMTI force plate in Live.


USB Access to AMTI Force Plates

To connect AMTI force plates via USB, first power on the amplifier, ensure the amplifier's power switch is turned on, and connect the force plate cable to the amplifier's "TRANSDUCER INPUT" port;

Connect one end of the synchronization cable to the AUX port at the back of the camera with IP 201, and the other end's white synchronization cable to the "GENLOCK/TRIGGER INPUT" port on the amplifier;

Insert the USB cable into the USB port on the amplifier, and connect the other end of the USB cable to the computer where the XINGYING software is installed;

Open the XINGYING software, open the Force Platform Configuration. In the "Other Configurations" at the bottom of the Force Platform Configuration window, check the "Digital Force Plate (USB)" option, click "+" and it will automatically add a force plate. Select the force plate name, and in the "Force Plate Specifications" on the right side of the window, set the type to AMTI, and set the length and width according to the size of the force plate; the parameters under "Orientation in Volume, Position in Volume" can use the system's default parameters, or you can modify them according to your needs. The "Serial ID" at the bottom of the window will display the serial number of the force plate. After configuring the force plate parameters, please click the "Apply" button to save this configuration (16.3.10);

Close the settings window, connect the camera, click the play button in the Live 3D view mode, at this point the AMTI force plate, which has been connected via USB, has been added to the 3D view and is ready to be used.

Last updated